    User name kuss

    Log entry time 22:04:10 on February28,2000

    Entry number 34971

    keyword=ITC502 controller reset

    Had to make an access to reset the ITC502_1 which contains the sensor that is used to PID the heaters. Apparently the ITC was operating, as could be checked on the video screen, but the connection was lost. IOC reboot didn't help, actually it resetted the sensor entries in the epics data base to 0.000, and the heaters tried to heat up the target. Switching off and on of the ITC solved the problem.

    Parasitically also resetted the HRSE angle camera movement.