• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name R. Michaels

    Log entry time 17:43:55 on December 12,1999

    Entry number 30881

    keyword=calibration of SPOT++ checked

    From home I looked at runs 1109 - 1113, movements
    of beam 2 mm up and down with HARP measurements.
    I checked calibration of SPOT++ code. Result: The
    plot has shows a spot too big by about 25% and a
    systematic shift of ~0.5 mm. This is not too dangerous
    and is correctable in offline analysis. I fixed the annoying
    problem of the sign of X, made it agree with EPICS.