Log entry time 08:16:28 on October 1,2000
Entry number 48859
keyword=spectrometer motion
mark stevens disconnected and isolated one of the motors in
the left arm which had a mechanical problem. now it should
work fine. however when it is being moved, one should set the
current limit (by hand in the GUI) to 10 A (from 5) in the middle
and outer bogies.
.
one will need to do this each time that the bogies program is restarted
since the GUI resets these limits to 5 A at restart.
.
none of this applies to the right arm motion.