• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name R. Michaels

    Log entry time 11:00:10 on October 1,2000

    Entry number 48882

    Followups:

    keyword=efficiency and VDC glitches

    Problems noted:
    1. VDCs on both spectrometers have a secondary TDC peak at
    large TDC values; it might be caused by after-pulsing on PMTs (?)
    See figure below.
    2. Trigger T2 is bizarre. It has lots of events with VDC hits but also
    lots event with neither S1 nor S2. Both T2 and T4 triggers need to
    be checked with a scope (it requires an access).
    3. Mark Jones noted a hole (inefficiency) in the momentum spectrum
    on right spectrometer; it seems to be around paddle 5 or 6.
    The ADC amplitudes look big enough, so its probably not a
    HV problem. He is studying it.


    FIGURE 1