    User name R. Michaels

    Log entry time 02:35:21 on October15,2000

    Entry number 50434

    keyword=HV on R-arm was off !

    Run 1863, we noticed that the T1 rate (R-arm trigger) was zero.
    The HV was turned off !! Probably was not off more than 1/2 hour.
    We turned it back on via portserver. Hmm... the EPICS GUI for
    ``Hall A HV Control'' was misleading - seems to indicate HV was on.
    Also, no EPICS readback of actual HV.